POLARIS

This page collects information that will be useful for the implementation of the IBEX control system on POLARIS.

Background & Timeline

POLARIS is a diffractometer instrument at ISIS, on TS1. The POLARIS web page describes the background to the instrument.

Control System

POLARIS will migrate from the SECI control system to the IBEX control system.

POLARIS Equipment

The equipment listed below is used on POLARIS. Please add new information (e.g. new items of equipment, new notes, information about drivers, etc. to this table).

Manufacturer

Model

Type

Connection

Driver

Notes

ISIS

DAE 2

Detector Electronics

Ethernet

see DAE note

ISIS

Mk2 Chopper

Chopper

RS-232

see Chopper note

GALIL

DMC2280

Motion Controller

Ethernet

EPICS

see Galil note

JJ-XRAY

4 x 4-blade jaws

Blades driven by Galils. see JJ X-Ray note

1 x 4-blade jaws

Blades driven by Galils. Mounted inside sample tank

ISIS

Mk3 Chopper

Chopper

Ethernet/.NET

#169

Pfeiffer

TPG300

ISIS Vacuum System

RS232

#216

see Pfeiffer note

Pfeiffer

TPG26x

Gas Handling System

RS232

#1411

see Pfeiffer note

Eurotherm

Temperature Controller

All models at ISIS

RS-232

EPICS

see Eurotherm note

Chell

CCD100

Pressure Transducer

RS-232

#1827

see Chell note

Keithley

2400

Source Meter

RS-232

#1826

see Keithley note

Oxford Instruments

Orange Cryostat

Cryogenic System

see Oxford Instruments note

Oxford Instruments

Blue Cryostat

Cryogenic System

see Oxford Instruments note

Oxford Instruments

Dilution Fridge (Kelvinox)

Cryogenic System

see Oxford Instruments note

McLennan

PM600

Motion Controller

RS-232

EPICS

see McLennan note

ISIS

Eurotherm & Keithley

Resistivity Measurements

RS-232

Only an offline experiment to date see Resistivity note

ISIS

POLARIS Sample Changer

Sample Changer

RS-232

Copy of GEM device

Applied Measurements Ltd

INT2-L

Pressure Transducer

RS-232

Used in the Hydrothermal Reaction Cell see Hydrothermal Reaction Cell note

ISIS

POLARIS Pressure Cell Controller

Pressure Cell Controller

RS-232

see Pressure Cell Controller note

Spellman

SL600

high-voltage power supply

N/A

see Spellman note

ISIS

CS155 centre stick

high voltage centre stick

N/A

see HV Centre Stick note

Note: DAE

Main Detector banks + fixed monitors.

Note: Choppers

One T0 chopper.
Chopper is ISIS Mk2 chopper. Mk2 choppers have a serial interface (not Ethernet like Mk3).

Note: JJ X-Ray

JJ X-Ray 4-blade jaws
All blades are driven by Galils.
POLARIS has two screens for managing all 5 sets of jaws:

  1. One screen to view & edit settings for jaw sets 1 -4 and display settings for jaw set 5

  2. A separate screen (Jaws Manager) to view & edit settings for all 5 jaw sets

Note: Pfeiffer

  1. Model TPG 26x

  2. Model TPG 300

Note: Chell

Experiments requiring pressure transducer. Chell CCD100

Note: Pressure Cell Controller

The pressure cell controller on POLARIS is a copy of the same device on PEARL. At the current time (April 2017), it has not been used on POLARIS for some time. However, it may be used at some future point (but probably not for another 12-18 months, by which time we should have implemented support for the pressure cell controller on PEARL).

Note: Keithley

Experiments requiring source meter. Keithley 2400 Source Meter

Note: Eurotherm

Used to control temperature Orange Cryostat devices.

Note: Oxford Instruments

  1. Orange Cryostat: models to be determined. Orange cryostats are controlled via a temperature controller (e.g. Eurotherm), so nothing on the cryostat itself for IBEX to control.

  2. Blue Cryostat: models to be determined.

    1. Are these the Heliox and ITC devices?

    2. The Heliox device has a sorption insert (is that relevant to the control system?)

  3. Dilution fridge: models to be determined.

    1. Kelvinox dilution fridge

Note: McLennan

  1. One McLennan is used raise/lower the sample-stick

  2. Another single-axis McLennan is used to rotate the sample-stick

Note: Galil

GALIL_02 is currently set up to use QR instead of DR mode because DR mode stopped working. This is the same issue as documented in Ticket #2191.

Note: Hydrothermal Cell

Uses the INT2-L pressure transducer.

Note: Resistivity Measurements

An evacuated furnace cell for measuring resistivity. Controlled by Eurotherms and measured by Keithleys (models 2000, 2400?)

Note: Spellman High Voltage Power Supply

6kV, 100mA, 600W, DC power supply. Not possible to control this device via software.

Note: CS155 Centre Stick

Not possible to control this device via software. It connects to the Spellman HV-PS.

POLARIS SECI Configs

Document information about POLARIS SECI configs here.

Configuration Name

Sub-Configurations

Last Accessed

Required

POLARIS Base + Hydrothermal

Chopper, Vacuum, Jaws, Eurotherms Test

03/02/2017

-

POLARIS Base + Dilution Fridge

Chopper, Vacuum, Jaws, Oxford - Dilution Fridge

15/12/2016

-

POLARIS Base + Sample Changer + Eurotherms_matt

Chopper, Vacuum, Jaws, Eurotherms Test, Sample Changer

12/12/2016

-

POLARIS Base + Sample Changer + Eurotherms + Heliox

Chopper, Vacuum, Jaws, Eurotherms Test, Sample Changer, Heliox

19/09/2016

-

POLARIS Base

Chopper, Vacuum, Jaws

30/11/2015

-

POLARIS Base + Sample Changer + Eurotherms

Chopper, Vacuum, Jaws, Eurotherms Test, Sample Changer

18/11/2015

-

POLARIS Base + Sample Changer + Eurotherms + ITC_matt

Chopper, Vacuum, Jaws, Eurotherms Test, Sample Changer

25/06/2015

-

POLARIS Base + Eurotherms

Chopper, Vacuum, Jaws, Eurotherms Test

05/06/2014

-

POLARIS Base + Keithley (Voltage) + Eurotherms

Chopper, Vacuum, Jaws, Eurotherms Test, Keithley (Voltage)

02/06/2014

-

POLARIS Base + Keithley + Eurotherms

Chopper, Vacuum, Jaws, Eurotherms Test, Keithley

25/03/2013

-

delete

-

01/10/2012

-

POLARIS Base + Sample Changer

Chopper, Vacuum, Jaws, Sample Changer

14/03/2012

-

test

-

10/11/2011

-

POLARIS Genie Scripts

Similarly, document information about SECI Genie scripts currently used on POLARIS here.

POLARIS Notes