NIMROD Jaws
Nimrod jaws are a set of 6 jaws which work in concert to limit the beam of neutrons reaching the sample (see Jaws Manager for more). The 6 jaws are ordered 1 nearest the beam and 6 nearest the sample.
The parameters, with reference to the diagram in Jaws Manager, are:
Letter |
Parameter |
Value |
Notes |
---|---|---|---|
c |
Collimator position |
0 mm |
No collimator so beam tapered from moderator |
j1 |
Jaw 1 position |
6458 mm |
Distance from moderator |
j2 |
Jaw 2 position |
8072 mm |
Distance from moderator |
j3 |
Jaw 3 position |
12072 mm |
Distance from moderator |
j4 |
Jaw 4 position |
15450 mm |
Distance from moderator |
j5 |
Jaw 5 position |
17235 mm |
Distance from moderator |
j6 |
Jaw 6 position |
19691 mm |
Distance from moderator |
s |
Sample position |
19691 mm |
Note that the sample is assumed to be at the final jawset |
w |
Width of moderator |
user set |
|
h |
Height of moderator |
user set |
|
x |
Width at sample |
user set |
|
y |
Height at sample |
user set |
Setup
The control files for the nimrod jaws are part of the galil motor set up and are per instrument. There is an example in ...\EPICS\support\motorExtensions\master\settings\nimrod_jaws\*.cmd
which would need to be copied to C:\Instrument\Settings\config\<instrument host name>\configurations\galil
. IOC System tests also exist for the logic.
Jaw centres
The jaws on NIMROD are not centred with respect to the beam, i.e. all jaws being at a centre of (0, 0) will not actually let the beam through. This is due to some of the jaw sets being mounted off-centre during installation.
The expected centres of each jaw set are:
Jaw set |
Horizontal |
Vertical |
---|---|---|
1 |
-0.135 |
-1.5 |
2 |
-0.126 |
-1.457 |
3 |
-0.205 |
-9.634 |
4 |
0.159 |
-1.460 |
5 |
-0.303 |
-0.801 |
6 |
-0.378 |
-9.507 |
The numbers above correspond to the jaws being centred around the beam in “real” coordinates as opposed to motor coordinates.