POLARIS Jaws

Polaris jaws are a set of 5 jaws which work in concert to limit the beam of neutrons reaching the sample (see Jaws Manager). Jaws 1-4 are instrument user controlled and work together in a pyramid shape to reduce the beam size in stages. The gap in each jaw set is set based on its distance from the sample. There are documents stored in sharepoint. The 5 jaws are ordered 1 nearest the beam and 5 nearest the sample. Jaw set 5 is under on instrument scientist control. To change these settings the user must put the GUI into management mode.

The parameters are (with reference to the diagram in Jaws Manager):

Letter

Parameter

Value

Notes

c

Collimator position

1610 mm

Distance from moderator

j1

Jaw 1 position

6502 mm

Distance from moderator

j2

Jaw 2 position

9440 mm

Distance from moderator

j3

Jaw 3 position

11085 mm

Distance from moderator

j4

Jaw 4 position

11735 mm

Distance from moderator

j5

Jaw 5 position

13175 mm

Distance from moderator

s

Sample position

14000 mm

Distance from moderator

w

Width of Collimator

82.2

h

Height of Collimator

79.4

x

Width at sample

user set

y

Height at sample

user set

Setup

The control files for the Polaris jaws are part of the galil motor set up and are per instrument. There is an example in ...\EPICS\support\motorExtensions\master\settings\polaris_jaws\*.cmd which would need to be copied to C:\Instrument\Settings\config\<instrument host name>\configurations\galil